Nyquist Calculator
Diffraction-limited resolution and Nyquist sampling from an objective’s magnification, NA and wavelength.
Pick an objective or type your own magnification and NA, set the wavelength, and see the diffraction-limited spot pair with its minimum resolvable distance and the Nyquist–Shannon sample pitch. Everything runs in your browser.
Objective
User defined
- Rayleigh d 0.61λ/NA
- –
- Abbe limit λ/2NA
- –
- Nyquist pitch d/2 at sample
- –
- Camera pixel ≤ d/2 × Mag
- –
Camera · does it sample at Nyquist?
- Camera pixel
- –
- Total magnification Mag × factor
- –
- Pixel at the sample
- –
- Nyquist needs ≤ d/2
- –
- Verdict
- –
Saved for comparison · overlaid on the profile above
| Objective | Mag / NA | λ | Rayleigh d | Nyquist d/2 | Camera | Px at sample | Nyquist? |
|---|
How it works
A diffraction-limited objective turns a point of light into an Airy pattern. Two points are “resolved” when they sit far enough apart that your detector can still see the dip between them.
- 1.
Rayleigh criterion
The classic minimum resolvable peak-to-peak distance puts the peak of one Airy pattern on the first dark ring of the other:
d = 0.61 λ / NA. At this spacing the midpoint dips to about 74% of the peaks — that is the valley in the cross profile. - 2.
Abbe limit
The hard diffraction limit for the finest grating an objective can pass:
d = λ / 2NA. It is slightly tighter than Rayleigh and is the number most deconvolution and super-resolution literature starts from. - 3.
Nyquist–Shannon sampling
To record detail of size
dwithout aliasing you need at least two samples across it, so the sample pitch at the specimen must be≤ d/2. The comb under the profile shows that pitch. - 4.
Camera pixels
The objective magnifies the image before the camera sees it, so the physical pixel may be
d/2 × Mag. A 60×/1.4 objective at 520 nm resolves 227 nm, so pixels up to about 6.8 µm sample it at Nyquist.
Good to know
- The objective list covers common Nikon, Olympus/Evident, Zeiss and Leica research objectives as a convenience. Specifications change between revisions — confirm against the maker’s catalog before you order anything.
- Use the emission wavelength for fluorescence. Beyond ~780 nm the slider shows near-infrared, which your eye (and this pseudocolor) barely sees.
- Rayleigh assumes an aberration-free, evenly illuminated pupil. Real systems resolve a little worse.
Matching a camera?
- Keep the physical pixel at or below d/2 × Mag. Oversampling is safe; undersampling aliases fine detail and no software brings it back.
- Questions about triggering or camera timing? That is what our Triggerscope is for.